Journal of International Reproductive Health/Family Planning ›› 2012, Vol. 31 ›› Issue (3): 217-219.

• 论著 • Previous Articles     Next Articles

Analysis on the Adverse Events of Copper-T IUD Breakage of 30 Cases

ZHOU Ding-jie,YAO Jie,WEI Yong-yue,LI Ying   

  1. Jiangsu Institute of Planned Parenthood Research,Nanjing 210036,China(ZHOU Ding-jie,YAO Jie,LI Ying);NCFPC Center for Contraceptives Adverse Reactor Surveillance,Nanjing 210036,China(ZHOU Ding-jie,YAO Jie,LI Ying);Department of Epidemiology & Biostatistics,School of Public Health,Nanjing Medical University,Nanjing 210029, China(WEI Yong-yue)
  • Received:1900-01-01 Revised:1900-01-01 Published:2012-05-15 Online:2012-05-15
  • Contact: LI Ying

Abstract: Objective: To evaluate the adverse events of the copper-T Intrauterine device(IUD)breakage, so as to explore the relative factors. Methods: A total of 30 cases of the breakage adverse events related to copper-T IUD were collected from 105 contraceptive adverse reaction surveillance sentinels from October 2005 to November 2010. All of these cases were analyzed by the method of causality assessment. Results: There were 4 TCu380A cases and 26 TCu220C cases in 30 cases. The fracture parts were cross arms in 28 cases and trailing arms in others. The breakage constituent ratio of TCu220C from B enterprise was higher than that from A enterprise(IC=0.40, lower limit of 95%CI=0.02). Conclusions: The postmarked surveillance of TCu220C from B enterprise should be strengthened. The professional staff of family planning should pay attention to the adverse events of copper-T IUD breakage. To prevent the serious injury events,it is key to keep up the follow-up visit system after insertion. The adverse events should be treated timely, and the serious injury event should be reported immediately.

Key words: Intrauterine devices, copper, Breakage, Adverse event, Serious injury event